ISSN 2734-245X
 

Original Article
Online Published: 20 Feb 2024
 


Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter

Kenneth O. Inaku, Lawson E. Ekpe, Asuquo B. Ene, Onuche L. Onuche, Essienawan U. Akpan, Lucy A. Inaku.


Abstract
Introduction: Frequent and accurate blood glucose measurements and control can delay the onset of diabetic complications. This is facilitated by the use of point-of-care glucose meters (POCGM). There is a need to verify some claims made by manufacturers of POCGM with regard to accuracy, precision and linearity. To verify some of these claims, we compared blood glucose results obtained from Accu-Check Instant blood glucose meters with laboratory results as a reference.
Method: This cross-sectional analytical study involved verifying some characteristics of AccuChek Instant POCGM using different concentrations of glucose calibrators prepared in-house. Blood was obtained from volunteers who did not need an overnight fast. It was tested using Accu-Chek Instant POCGM and routine laboratory glucose oxidase method as a reference and compared using the Bland-Altman plot. Statistical significance was set at p<0.05.
Results: There were 188 blood samples tested using the two methods. The difference between the means of glucose values from the two methods was 10.3 mg/dL (12.8%), with Accu-Chek POCGM returning higher results than the routine laboratory method. The difference was not statistically significant using the Bland-Altman plot (p=0.60). The analytical range of the POCGM was determined to be between 20mg/dl to 300 mg/dL. Results indicating ‘Hi’ on the POCGM were below 600 mg/dL, contrary to what was claimed by the manufacturer.
Conclusions and Implications: There was some agreement between the two methods studied. Results obtained at high glycaemic states may be more precise but are lower than central laboratory values.

Key words: Diabetes mellitus type 2, Blood glucose self-monitoring, Fasting blood glucose, Reproducibility of results, Humans


 
ARTICLE TOOLS
Abstract
PDF Fulltext
How to cite this articleHow to cite this article
Citation Tools
Related Records
 Articles by Kenneth O. Inaku
Articles by Lawson E. Ekpe
Articles by Asuquo B. Ene
Articles by Onuche L. Onuche
Articles by Essienawan U. Akpan
Articles by Lucy A. Inaku
on Google
on Google Scholar


How to Cite this Article
Pubmed Style

Inaku KO, Ekpe LE, Ene AB, Onuche OL, Akpan EU, Inaku LA. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. crjmed. 2024; 3(1): 8-12. doi:10.5455/CRJMED.175560


Web Style

Inaku KO, Ekpe LE, Ene AB, Onuche OL, Akpan EU, Inaku LA. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. https://www.crjmed.com/?mno=175560 [Access: October 28, 2024]. doi:10.5455/CRJMED.175560


AMA (American Medical Association) Style

Inaku KO, Ekpe LE, Ene AB, Onuche OL, Akpan EU, Inaku LA. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. crjmed. 2024; 3(1): 8-12. doi:10.5455/CRJMED.175560



Vancouver/ICMJE Style

Inaku KO, Ekpe LE, Ene AB, Onuche OL, Akpan EU, Inaku LA. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. crjmed. (2024), [cited October 28, 2024]; 3(1): 8-12. doi:10.5455/CRJMED.175560



Harvard Style

Inaku, K. O., Ekpe, . L. E., Ene, . A. B., Onuche, . O. L., Akpan, . E. U. & Inaku, . L. A. (2024) Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. crjmed, 3 (1), 8-12. doi:10.5455/CRJMED.175560



Turabian Style

Inaku, Kenneth O., Lawson E. Ekpe, Asuquo B. Ene, Onuche L. Onuche, Essienawan U. Akpan, and Lucy A. Inaku. 2024. Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. Cross River Journal of Medicine, 3 (1), 8-12. doi:10.5455/CRJMED.175560



Chicago Style

Inaku, Kenneth O., Lawson E. Ekpe, Asuquo B. Ene, Onuche L. Onuche, Essienawan U. Akpan, and Lucy A. Inaku. "Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter." Cross River Journal of Medicine 3 (2024), 8-12. doi:10.5455/CRJMED.175560



MLA (The Modern Language Association) Style

Inaku, Kenneth O., Lawson E. Ekpe, Asuquo B. Ene, Onuche L. Onuche, Essienawan U. Akpan, and Lucy A. Inaku. "Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter." Cross River Journal of Medicine 3.1 (2024), 8-12. Print. doi:10.5455/CRJMED.175560



APA (American Psychological Association) Style

Inaku, K. O., Ekpe, . L. E., Ene, . A. B., Onuche, . O. L., Akpan, . E. U. & Inaku, . L. A. (2024) Comparison of Methods and Verification of Some Performance Characteristics of Accu-Chek® Instant Glucose Meter. Cross River Journal of Medicine, 3 (1), 8-12. doi:10.5455/CRJMED.175560